Keisuke Yasui   Toshiyuki Toshito   Chihiro Omachi   Kensuke Hayashi   Hideto Kinou   Masaki Katsurada   Naoki Hayashi   Hiroyuki Ogino   
Journal of applied clinical medical physics 20(2) 114-120 2019年2月 [査読有り]
The purpose of this study was to propose a verification method and results of intensity-modulated proton therapy (IMPT), using a commercially available heterogeneous phantom. We used a simple simulated head and neck and prostate phantom. An ioniza...