Cell Disruption Device
Contents on page
Cooling Bead-Based Cell Disruption Device
| Model (Manufacturer) |
MicroSmashMS-100R (TOMY) |
| Features and Specifications, etc. | [Overview] This equipment can crush bacteria, yeast, animal tissue, plant tissue, etc. under cooling (bead-type crusher) . [Specifications] - Capacity: 2mL tube x12 - RPM setting range: 2,000 to 5,500 rpm (in 100 rpm increments) - Crushing control method: Vertical rotation 3D high-speed motion method - Cooling temperature: 2°C in the crushing chamber for about 15 minutes - Mode: Manual mode, memory mode (5 patterns) - Timer setting range: 1 to 300 seconds (up to 100 seconds for 5,100 rpm and above) - Safety devices: door open/close detection, motor heating protection circuit, power knob looseness detection - Display method: LCD digital display (with backlight) Beads: glass beads, zirconia beads, zirconia/silica beads, stainless steel beads, tungsten beads [Features] - Effective crushing is achieved by the 3D high-speed motion of the tube, which swivels up and down and in a figure-8 pattern, to increase the collision between the sample and the beads. - Transparent door allows observation of the inside of the crushing chamber from the outside. - It can also emulsify samples. |
| Location | Central Research Center (Room 317, University Building 1) |
| Remarks | - New users are requested to contact the Cell Function Analysis Laboratory. - Please enter your reservation through the reservation system before using. - Please limit the use (reservation) time to up to 3 hours. |
| Equipment Manager Extension Number |
Tezuka (Cell Function Analysis Laboratory) 2570 |
| Usage Fee | 100 yen/run |
Equipment Photo
Preparation.