Liquid Chromatography System
Contents on page
High Performance Liquid Chromatography System
| Model (Manufacturer) |
Nexera Series (SHIMADZU) |
| Features and Specifications, etc. | An instrument that separates compounds dissolved in a liquid sample and performs qualitative and quantitative analysis to determine what components are contained and how much they are contained. In HPLC analysis, the liquid that flows through the components in a liquid sample for separation is called the mobile phase. The mobile phase flows at a constant rate from the column (stationary phase) to the detector by means of a pump. A sample containing a mixture of multiple components is injected into the column and separated in the column. The components are monitored by a detector installed at the column outlet for qualitative and quantitative analysis. In addition to an absorbance detector that measures at wavelengths in the UV and visible regions, this system is equipped with a fluorescence detector that measures compounds that emit fluorescence. Please prepare mobile phase and columns by yourself. Approximate column size: 2-3 μm particle size, 2-4.6 mm i.d. 【Specifications】 Pumping unit LC-40DXR: maximum allowable pressure 70 MPa Detector: Absorbance detector SPD-40V: wavelength range 190-1000 nm Fluorescence detector RF-20A XS: wavelength range 200-750 nm Autosampler SIL-40C XR: Number of samples processed: 162 1.5 mL reagent bottles Injection volume range 0.1-50 μL |
| Location | Omics Analysis Laboratory (Room 304, University Building 1) |
| Remarks | - |
| Equipment Manager Extension Number |
Maeda (Omics Analysis Laboratory) 9814 |
| Usage Fee | 100 yen/hour *Figures less than one hour are rounded up. |