Skip global navigation and go to header navigation

Skip global navigation and read the article

Skip global navigation and go to local navigation

Skip global navigation and go to footer navigation



TOP >  Installed Equipment and Facilities >  Metabolic Analyzer

Metabolic Analyzer


Extracellular Flux Analyzer

Model
(Manufacturer)
XFe24 (Agilent Technologies)
Features and Specifications, etc. [Features]
- It is an analyzer that automatically analyzes cellular metabolism, allowing non-invasive and highly sensitive temporal measurements of major energy metabolism pathways such as glycolysis and aerobic respiration by mitochondria. It finds applications in various research fields including cancer, immunology, diabetes, obesity, neurodegenerative diseases, and cardiovascular diseases.

- Simultaneous measurement of two cellular energy metabolisms: mitochondrial respiration ≒ oxygen consumption rate (OCR) and glycolysis ≒ extracellular acidification rate (ECAR)

- Detecting the changes in oxygen concentration (pmol/min) and hydrogen ion concentration (mpH/min) in cultured cells or isolated mitochondria using luminescence methods, and calculate OCR and ECAR .

- High-throughput and rapid automated measurements:
Measurements are performed automatically by culturing cells on dedicated 24-well (XFe24 model) simply setting them into the main unit.

[Specifications]
- Analytical method: Photoluminescence method
- Simultaneous measurement parameters:
O2 ・H+
- Number of simultaneous measurement samples: 24
- Number of injection ports: 4
- Injection port volume: 75 µL
- Optimal measurement range for OCR (pmol/min) : 50 to 400
- Optimal measurement range for ECAR (mpH/min) : 20 to 120

[URL]
https://www.primetech.co.jp/products/tabid/90/pdid/36/language/ja-JP/Default.aspx
Location Central Research Center (Room 317, University Building 1)
Remarks - This equipment can be used after attending a training session conducted by the staff.
Please contact the person in charge regarding the training.
- When using this device, please be sure to reserve it through the equipment reservation system.
- Please be sure to fill in the usage log provided on the equipment whenever you use it.
Equipment Manager
Extension Number
Sugihara (Laboratory of Genome Sequencing and Analysis)
9372
Usage Fee 500 yen/time
Equipment Photo